PIP - QUIK-COMBO Defibrillator Delivered Energy Tests at 2J and 70 J
To perform QUIK- COMBO Defibrillator Delivered Energy Test at 2J and 70J:
1. Set the Defibrillator Analyzer to measure
ENRG
2. Press
ENERGY SELECT
on the device and select
2 J.
3. Press the
CHARGE
button and wait for the device to reach full charge. Then push the
SHOCK
button to discharge the device.
4. Verify the defibrillator analyzer indicates the delivered energy is within 1.0 J to 3.0J.
5. Repeat steps 2 and 3 for the energy level at 70J.
6. Verify the defibrillator analyzer indicates the delivered energy is within 59.5 J to 80.5J.
Note:
Perform the
TCP- Defibrillator Energy Calibration
if the delivered energy falls outside of the acceptable output range.