Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 32 of 61
PIP - Standard Paddles User Test
Perform this test if STANDARD paddles are used. Otherwise, skip to the next test.
Note:
Use the customer’s standard paddles (when available). Remove the paddles and check that the paddle surfaces and paddle wells
are clean and dry and free of any debris. Verify that the metal surface of the standard paddles and paddle test contacts in the device
paddle wells are free of burn and arc marks. Also check that these surfaces are free of pits, scratches or raised nicks that can be felt
with the finger tip. Check the therapy connector interface for pin damage.
WARNING
SHOCK HAZARD
The conductive gel (wet or dry) on the paddle handles and in the paddle wells may allow the electrical energy to arc between
paddles during discharge. Thoroughly clean and dry the paddles and the paddle wells after use and before performing the
Standard Paddles User Test.
1. Connect the standard paddles to the device. Place the paddles in the
paddle wells as shown in Figure 1.20.
2. Turn the device
ON
.
3. Set the device Lead selection to
PADDLES
.
4. Press the
OPTIONS
button and select
USER TEST
from the Option
screen.
5. Select
YES
from the Option/User Test screen.
6. Push Speed Dial to initialize the Self-Test and the User Test. The Self-
Test and the User Test are performed. The User Test Succeeded
report is printed when test is complete.
Note:
The Device will
automatically turn off after successfully completing the test.
7. Proceed to next test.
Figure 1.20: Standard Paddles User Test Setup