LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 42 of 61
PIP - Pacer Output Current Test
To perform the pacer output current:
Note:
Perform the pacer output current test at 10 mA, 100 mA and 200 mA. You must repeat the test for each current level.
1. Select Pacer button on Impulse 7000DP to measure pacing current.
2. In the menu screen, set the Brand to "Physio-Control" Input Jacks to "Defib," and Load to 50 ohms.
3. Set Pacer Rate at 60PPM.
4. Press CURRENT on the device, and select a pacer current (10 mA, then 100 mA, and then 200 mA).
5. Verify the defibrillator analyzer indicates the pacer output current is within the acceptable output limits shown in Table below (results
may appear as negative numbers).
Note:
If the Impulse 7000DP does not detect a pacing output current reading, then operate the Impulse 7000DP and device on
battery power for the Pacer output current test.
Note:
Perform TCP
– Pacer Self-Calibration if the peak pacer current falls outside the acceptable output range.
Peak Current Level (mA)
Acceptable Output (mA)
10
0 to 20
100
90 to 110
200
180 to 220