LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 36 of 61
PIP - Standard Paddles Defibrillator Delivered Energy and Charge Time Tests at 360J (Battery Powered)
WARNING
SHOCK HAZARD
Electrical energy is discharged during this procedure. Do not allow the electrodes to contact any person or conductive surface
except as described below.
To perform Standard Paddles Defibrillator Delivered Energy and Charge Time Tests at 360J (Battery Powered)
1. Establish the setup as shown in Figure 1.22.
2. Disconnect the AC power cable from the device and turn the
device
ON
by battery.
3. Press the
ENERGY SELECT
button to select 360 J.
4. Start a Stop Watch, then press the
CHARGE
button on Standard
Paddles and wait for the device to reach full charge. Stop a Stop Watch,
and then push both
SHOCK
buttons on Standard Paddles to discharge
the device.
5. Verify the defibrillator analyzer indicates the delivered energy is within
306.0 J to 414.0 J.
6. Verify the Charge Time is within 1s to 10 s.
Note:
Perform the TCP- Defibrillator Energy Calibration if the delivered
energy falls outside of the acceptable output range.
7. After testing is complete reconnect the AC power cable to the device.
Figure 1.22: Standard Paddles Defibrillator
Energy and Charge Time Test Setup