LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 20 of 61
Note:
Perform this test if 3-lead ECG cable is used. Otherwise, skip to the next test.
The 3-Lead ECG tests consist of:
PIP - 3-Lead ECG Leads Off Detection Test
PIP - 3-Lead ECG Gain Test
PIP - 3-Lead ECG Leads Off Detection Test
1.
Connect the 3-Lead ECG cable between the device and Impulse 7000DP as
shown in Figure 1.10.
2.
Set the Impulse 7000DP output to a 1-mv, 10-Hz sine wave.
3.
Set the device
LEAD
selection to
Lead II
.
4.
Remove the
LA lead
from the Impulse 7000DP, and verify the device displays
the
ECG LEADS OFF
message and a repeating priority 3 tone shall sound when
the Lead is removed. Reconnect the LA lead.
5.
Remove the
RA lead
from the Impulse 7000DP, and verify the device displays
the
RA LEADS OFF
message and a repeating priority 3 tone shall sound when
the Lead is removed. Reconnect the RA lead.
6.
Remove the
LL lead
from the Impulse 7000DP, and verify the device displays
the
LL LEADS OFF
message and a repeating priority 3 tone shall sound when
the Lead is removed. Reconnect the LL lead.
7.
Continue with the next test with this setup in place.
Figure 1.10: 3 Lead ECG Leads Test Setup