LIFEPAK
®
20
Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 58 of 61
PIP-Earth Leakage Test - Normal Condition (NC)
WARNING
POSSIBLE EQUIPMENT DAMAGE
Do not defibrillate when connected to the safety analyzer!
1. Turn the device ON.
2. Press the
A button on the Safety Analyzer.
3. Press the F1/EARTH button on the Safety Analyzer and set the Safety Analyzer controls as follows:
Note:
Pause briefly between switching polarity to prevent damage to the defibrillator.
4. Verify the device AC Mains LED is ON.
5. Verify the measured current is between 15
A and 2250
A (120 VAC and 240 VAC).