Performance Inspection Procedure (PIP)
3201896-000_B
©2019 Stryker
Page 18 of 61
Printer Verification @ 25 mm/sec
1. Select
PRINTER
in the
Service/Tests
submenu.
2. Select
START
to print a test strip.
3. Inspect the test strip for the following attributes:
a.
The large “X” form prints without missing dots.
b. Four horizontal lines print (one very close to the lower paper margin).
c. The character set prints clearly without broken characters.
d. Vertical lines spaced 25 mm ±1mm (approx. 24 to 26 mm.) apart print
correctly.
4. Open the printer door and verify the
CHECK PRINTER
message appears at
the bottom of the screen.
5. Remove the printer paper, and then close the printer door.
6. Verify the
CHECK PRINTER
message appears at the bottom of the screen.
7. Select
PREVIOUS PAGE
twice to return to the Service menu.
8. Continue with the next test.
Figure 1.9: Printer Test