Oximeter

Clinician’s Operation Manual

Catalog Number 1850

Version 15, March 2005

© 2005 Smiths Medical family of companies.

All rights reserved.

Содержание 1850

Страница 1: ...Oximeter Clinician s Operation Manual Catalog Number 1850 Version 15 March 2005 2005 Smiths Medical family of companies All rights reserved ...

Страница 2: ......

Страница 3: ...ted Smiths family copyright statement Updated registered trademark statements Changed font to Myriad Pro Added this Revision History page Updated the Smiths logo to include a symbol Smiths and BCI and Comfort Clip are trademarks of the Smiths Medical family of companies Microfoam is a trademark of the 3M Company The symbol indicates the trademark is registered in the U S Patent and Trademark Offic...

Страница 4: ......

Страница 5: ...Chapter 2 Monitor Features 2 1 About the Manuals 2 1 Theory of Operation 2 1 Intended Use 2 2 Monitor Features 2 2 Chapter 3 Controls 3 1 Chapter 4 Operating Instructions 4 1 Unpacking the Monitor 4 1 AC Power 4 1 Attaching the Sensor to the Patient 4 2 Choosing the Sensor 4 3 Care and Handling of the Sensor 4 4 Checking the Sensor and Oximetry Cable 4 6 Cleaning or Disinfecting the Sensors 4 10 T...

Страница 6: ...or Trend Data 7 2 Compatible Printers 7 3 What You ll Need for Printing 7 3 Setting Up the Monitor and the Printer 7 4 Trend Printouts 7 5 Collecting Trend Data 7 5 Manually Incrementing the Patient Number 7 5 Clearing Trend Data 7 5 Printing Trend Data 7 6 Chapter 8 Operating Modes 8 1 About the Monitor s Operating Modes 8 1 Home Use Mode 8 1 Setting Up the Monitor for Home Use 8 2 Equipment and ...

Страница 7: ...anual iii Chapter 10 Maintenance 10 1 Schedule of Maintenance 10 1 Correcting the SENSOR Alert 10 2 Chapter 11 Troubleshooting 11 1 Chapter 12 Optional Supplies and Accessories 12 1 Ordering Information 12 2 Chapter 13 Specifications 13 1 Oximeter 13 1 ...

Страница 8: ...This page is intentionally left blank ...

Страница 9: ...arding any specification Warranty Limited Warranty Seller warrants to the original purchaser that the Product not including accessories shall be free from defects in material and workmanship under normal use if used in accordance with its labeling for two years from the date of shipment to the original purchaser Seller warrants to the original purchaser that the reusable oximeter sen sors supplied...

Страница 10: ...not properly maintained or recharged or repaired by persons not authorized by Seller Misuse includes but is not limited to use not in compliance with the labeling or use with accesso ries not manufactured by Seller This warranty does not cover normal wear and tear and maintenance items Limitation of Remedies The original purchaser s exclusive remedy shall be at Seller s sole option the repair or r...

Страница 11: ...he USA contact your local distributor Keep all original packing material including foam inserts If you need to ship the device use only the original packaging material including inserts Box and inserts should be in original condition If original ship ping material in good condition is not available it should be purchased from Smiths Medical PM Inc Damages occurred in transit in other than original...

Страница 12: ...ion 6 Type B equipment 2 Attention consult accompanying documents 7 Refer servicing to qualified service personnel B Alarm Silence 1 Non AP Device O On P Off J ID Clear Up and Down Keys I Alarm Volume K Pulse Volume C Use By REF Catalog Number y Date of Manufacture SN Serial Number K Moisture Sensitive ...

Страница 13: ...ice by or on the order of a physician 2 WARNING Do not use this device in the presence of flammable anesthetics 2 WARNING Do not use this device in the presence of magnetic resonance imaging MR or MRI equipment 2 WARNING This device must be used in conjunction with clini cal signs and symptoms This device is only intended to be an adjunct in patient assessment 2 WARNING Prolonged use or the patien...

Страница 14: ...e approved by Smiths Medical PM Inc 2 WARNING When connecting this monitor to any instrument verify proper operation before clinical use Refer to the instru ment s user manual for full instructions Accessory equipment connected to the monitor s data interface must be certified according to the respective IEC standards i e IEC 950 for data processing equipment or IEC 601 1 for electromedical equip ...

Страница 15: ...roved equipment must be placed out side of the patient environment The patient environment is defined as an area 1 5 m 4 92 feet from the patient Example of the patient environment The dimensions are not prescriptive Figure 1 1 Patient Environment side Figure 1 2 Patient Environment top ...

Страница 16: ... monitor voids the warranty CAUTION This device is intended for use by trained healthcare professionals The operator must be thoroughly familiar with the information in this manual before using the device CAUTION Monitors not in use must be fully charged at least every five months otherwise the battery will no longer be able to accept a charge CAUTION Connect only the printer cable specifically in...

Страница 17: ...cein may adversely affect the accuracy of the SpO2 reading NOTE Any condition that restricts blood flow such as use of a blood pressure cuff or extremes in systemic vascular resis tance may cause an inability to determine accurate SpO2 and pulse rate readings NOTE Remove fingernail polish or false fingernails before applying SpO2 sensors Fingernail polish or false fingernails may cause inaccurate ...

Страница 18: ...This page is intentionally left blank ...

Страница 19: ...ximeter determines SpO2 and pulse rate by passing two wave lengths of light one red and one infrared through body tissue to a pho todetector During measurement the signal strength resulting from each light source depends on the color and thickness of the body tissue the sensor placement the intensity of the light sources and the absorp tion of the arterial and venous blood including the time varyi...

Страница 20: ... home environment Finally Sleep Study Mode allows the health care professional to record sleep study data which may later be transferred to a PC and analyzed Monitor Features Provides fast reliable SpO2 pulse rate and pulse strength measure ments on any patient from neonates to adults Ideally suited for use in intensive care units outpatient clinics emer gency rooms during emergency air or land tr...

Страница 21: ...es of trend data in a graphic format SpO2 and pulse rate averaging settings are user selectable Trend view allows operator to view previous patient history Home use mode allows a home use caregiver to monitor a patient at home Sleep study mode allows the health care professional to record the SpO2 and Heart Rate values of a patient every 4 seconds for up to 12 hours Sleep study mode allows the hea...

Страница 22: ...This page is intentionally left blank ...

Страница 23: ...n oximetry cable can be connected between the monitor and the sensor The printer is also connected here 2 SpO2 Numeric Display A number shows the patient s SpO2 value in percent Dashes mean the monitor is not able to calculate the SpO2 value 3 Pulse Rate Numeric Display A number shows the patient s pulse rate value in beats per minute Dashes mean the monitor is not able to calculate the pulse rate...

Страница 24: ...sensor is not properly attached to the patient 2 WARNING While SENSOR is flashing the monitor cannot measure the patient s SpO2 or pulse rate You must immediately check the patient s condition After you have checked the patient s condition you must correct the SENSOR alert 7 BATT Light BATT flashes on and off when about 30 minutes of battery use remains The monitor will work until the battery beco...

Страница 25: ...ced light flashes on and off when the alarm and alert tones are silenced for two minutes The alarm silenced light remains on when the alarm and alert tones are silenced indefinitely until canceled or until the monitor is turned off 9 POWER Light POWER light is green when the power supply is attached 10 CHGING Light The CHGING light is yellow when the battery is fast charging ...

Страница 26: ...in the SpO2 digits Pressing and holding I D CLEAR for about six seconds clears all trend data and sets the patient number to 1 While the sensor is not connected to the monitor Pressing I D CLEAR causes the monitor to enter the trend view mode two bar graph seg ments flash to indicate trend view mode During trend view mode the last valid measurement for each patient number can be shown by press ing...

Страница 27: ...er Clinician s Operation Manual 3 5 14 and Keys The and keys are used to adjust up and down the following set tings brightness of the display alarm limits trend view patient num bers SpO2 and pulse rate averaging 15 ALARM SEL Key Pressing ALARM SEL cycles through each of the alarm limits for setting ...

Страница 28: ...alarm tone for two minutes Pressing and holding the alarm silence key for about three seconds silences the alarm tone indefinitely until canceled or until the monitor is turned off 17 ALARM VOL Key Pressing ALARM VOL key changes the alarm volume from soft to loud or loud to soft 18 PULSE VOL Key Pressing the PULSE VOL key changes the pulse beep volume NOTE The pulse volume is stored after the moni...

Страница 29: ... Power Supply Chapter 3 Controls Oximeter Clinician s Operation Manual 3 7 19 Sensor Printer Connector An optional printer can be connected for printing trend data See Printer section for more printer options 20 AC Power Supply AC power supply connects here ...

Страница 30: ...This page is intentionally left blank ...

Страница 31: ...using this monitor The monitor s AC rating is shown on the external power supply If the rating is not correct do not use the monitor contact the Smiths Medical PM Inc service department or your local dis tributor for help Refer also to chapter 12 Optional Supplies and Accessories to verify the proper AC power supply for your application NOTE Do not plug the monitor into an outlet controlled by a w...

Страница 32: ...inaccurate readings and blisters on the patient s skin lack of skin respiration not heat causes the blisters Attaching the patient to the monitor requires these steps 1 Choose the sensor 2 Check the sensor and oximetry cable 3 Clean or disinfect the sensor if using the reusable type Disposable sensors are for single patient use and do not require cleaning or dis infecting 4 Attach the sensor to th...

Страница 33: ... 3043 Sensor Universal Y 1300 Sensor Disposable Adult Finger Ear 3078 Sensor Ear Pediatric 15 45 Kg Finger 3044 Sensor Reusable Adult 3444 Sensor Reusable Comfort Clip 3178 Sensor Reusable Pediatric Finger or Toe 3043 Sensor Universal Y 1301 Sensor Disposable Pediatric Finger Ear 3078 Sensor Ear Infant 3 15 Kg Hand or Foot Toe Finger or Toe 3043 Sensor Universal Y 3025 Sensor Wrap Infant 1303 Sens...

Страница 34: ...d result in damaging of the sensor This may cause inaccurate readings Hold the connector rather than the cable when connecting or discon necting the finger sensor to the device as shown in Figure 4 1 Figure 4 1 Hold the Connector Connector Sensor Patient Cable Do not use excessive force or unnecessary twisting when connecting disconnecting storing or when using the sensor ...

Страница 35: ...sensor on the patient allow the cable to lay across the palm of the hand and parallel to the arm of the patient as shown in Fig ure 4 2 Figure 4 2 Finger Sensor Sensor Cable Upon completion of patient monitoring detach the sensor as shown in Figure 4 1 and loosely coil the finger sensor cable ...

Страница 36: ... readings Inspect each sensor If a sensor appears damaged do not use it Use another sensor or contact your authorized repair center for help 2 WARNING Using a damaged oximetry cable may cause inaccu rate readings Inspect the oximetry cable If the oximetry cable appears damaged do not use it Contact your authorized repair center for help 1 Carefully inspect the sensor to make sure it does not appea...

Страница 37: ...3 If using the oximetry cable a If the sensor is not already connected to the oximetry cable connect the sensor to the oximetry cable as shown Push the connectors together firmly and close the latch to secure the connectors b If the oximetry cable is not already connected to the moni tor connect the oximetry cable to the monitor as shown Push the connector firmly into the monitor ...

Страница 38: ...perating Instructions 4 8 Oximeter Clinician s Operation Manual 4 If not using the oximetry cable Connect the sensor to the monitor as shown Push the connector firmly into the monitor 5 If the monitor is not already on press the ON key to turn on the mon itor ...

Страница 39: ...se another sensor or oximetry cable or contact the equipment dealer for help if necessary Figure 4 5 Checking the Integrity of the Sensor Oximetry Cable and Oximeter Sensor not yet attached to patient SENSOR Indicator a Make sure the red light in the sensor is illuminated b Make sure the SENSOR indicator is flashing as follows NOTE Obstructions or dirt on the sensor s red light or detector may cau...

Страница 40: ...the patient Figure 4 6 Attach the finger sensor Attach the finger sensor to the patient as shown Be sure to fully insert the patient s finger into the sensor For patients with long fingernails use the universal Y sensor Cleaning or Disinfecting the Sensors Clean or disinfect reusable sensors before attaching to a new patient 2 WARNING Do not autoclave ethylene oxide sterilize or immerse the sensor...

Страница 41: ... at a time The monitor s software revision is momentarily displayed The patient number is momentarily displayed The monitor is in Clinician Mode Figure 4 7 SpO2 Pulse Rate and Pulse Strength Bar Graph Patient s SpO2 Shown Here Patient s Pulse Rate Shown Here Patient s Pulse Strength Shown Here 2 After a few seconds the SpO2 value pulse rate and pulse strength bar graph should be shown If not see t...

Страница 42: ... setting press and hold the arrow key while turning on the monitor While holding the key press the key to scroll through the following averaging selections SPO2 AVERAGING PULSE AVERAGING 4 8 8 8 16 8 16 16 NOTE SpO2 averaging is the number of pulse beats over which the SpO2 value is averaged pulse averaging is the number of seconds over which the pulse value is averaged ...

Страница 43: ...e Figure 4 8 Alarm Example SpO2 numbers flash during an SpO2 alarm Pulse Rate numbers flash dur ing a pulse rate alarm Alarm tone will sound if the Alarm Silenced light is off Alarm tone will not sound if the Alarm Silenced light flashes or is on During an alarm the numbers flash that correspond to the alarm the alarm tone sounds if not silenced The alarm tone sounds like a siren dee doo dee doo N...

Страница 44: ... pulse rate Sensor light flashes on and off Alert tone will sound if the Alarm Silenced light is off Alert tone will not sound if the Alarm Silenced light flashes or is on During an alert the SENSOR light flashes the alert tone sounds if not silenced The alert tone is a single tone sound with a pause beep beep pause beep beep 2 WARNING While SENSOR is flashing the monitor cannot measure the patien...

Страница 45: ...dur ing a BATT Attention During the BATT attention a short burst of beeps sounds every 30 seconds the BATT light flashes on and off 2 WARNING When BATT flashes you must immediately charge the monitor s battery Otherwise the monitor turns itself off about 30 minutes after BATT begins to flash See BATT atten tion in the Troubleshooting section for help ...

Страница 46: ...ired an optional Oximeter ECG Patient Simulator Smiths Medical PM catalog number 1606 is available as an accessory The simulator attaches to the oximeter in place of the sensor or oximetry cable It provides a known SpO2 and pulse rate signal to the oximeter This allows the oximeter s performance to be checked NOTE The 1606 Oximeter ECG Patient Simulator does not calibrate the monitor the monitor d...

Страница 47: ... about three seconds The alarm silenced indica tor lights steady while alarms are silenced indefinitely 3 To cancel either the indefinite or the two minute alarm and alert tone silenced condition momentarily press alarm silence key the alarm silenced indicator turns off Changing the Alarm and Alert Tone Volume The alarm and alert tones sound at one of two volumes soft or loud To change between sof...

Страница 48: ...o set the volume to loud Changing the Alarm Limits Each measurement SpO2 and Rate has a high and low alarm limit set ting Press the ALARM SEL key until the alarm limit you want to change is shown then press the or key to increase or decrease the setting ALARM SEL KEY PRESS DISPLAY ALARM LIMIT First press KI High SpO2 alarm limit Example only Second press 85 Lo 85 Low SpO2 alarm limit Example only ...

Страница 49: ...g mode is exited and the SpO2 and pulse rate measurements are shown NOTE Alarms are not active while setting alarm limits how ever alarms are active as soon as you exit the alarm limit set ting mode NOTE The low SpO2 alarm limit minimum test value is 80 If an operator changes the low SpO2 alarm limit to a value less than 80 and a power down power up sequence takes place a mini mum value of 85 take...

Страница 50: ...This page is intentionally left blank ...

Страница 51: ...is displayed only and is not incremented The monitor remembers all the trend data and all the patient numbers for up to 99 patients and 24 hours of run time Trend data for any patient can be viewed on the monitor Trend data for all patients can be printed on the optional printer NOTE See Printer section for information on printing trend data Manually Incrementing the Patient Number 1 The SpO2 sens...

Страница 52: ...e the patient number the patient number will be shown for a few seconds then the last valid data corresponding to the patient number will be shown e To exit the trend view mode press the I D CLEAR key NOTE If no keys are pressed for 20 seconds the monitor auto matically exits the trend view mode Clearing Trend Data 1 The SpO2 sensor must be connected to the monitor If necessary connect the SpO2 se...

Страница 53: ...up to 24 hours of previously stored data collected from 1 to 99 patients is printed Whenever the monitor is on it stores one SpO2 and pulse rate reading every thirty 30 seconds If less than 15 minutes of data is collected for one patient the trend data is printed in a tabular format if 123 mode is selected If more than 15 minutes of data is collected for one patient the trend data is printed in a ...

Страница 54: ...Trend Data 1 Keep the ALARM SEL key pressed for 5 seconds 123 or lll appears in the SpO2 display 2 123 numerical print out 3 lll graphic print out only possible with Seiko DPU 201 GSU or Seiko DFU 411 S printers 4 Select mode with arrow key The selec tion remains stored and will apply when the monitor is next switched on 5 To return to measurement press ALARM SEL key once If a key is not pressed w...

Страница 55: ... 21BU 120 V 60 Hz DPU 411 21BE 220 V 50 Hz NOTE A DB 9 to DB 25 pin adapter will be required For information on printers and printer accessories contact your autho rized sales representative What You ll Need for Printing You ll need these items to print trend printouts Oximeter Printer Cable see the Optional Supplies and Accessories section for ordering information Compatible printer purchased fro...

Страница 56: ...op bit no par ity 2 Connect the printer adapter s connector to the sensor connector 3 Connect the printer adapter s DB 9 connector to the mating connec tor on the printer If necessary connect the printer adapter s DB 9 connector to a DB 9 to DB 25 adapter then connect the adapter to the printer 4 Connect the printer s power source to the printer as described in the printer s operation manual CAUTI...

Страница 57: ...s patient the patient number is displayed only and is not incremented The monitor remembers all the trend data and all the patient numbers for up to 99 patients and 24 hours of run time Manually Incrementing the Patient Number 1 The SpO2 sensor must be connected to the monitor If necessary connect the SpO2 sensor to the monitor 2 Press the I D CLEAR key to increment the patient number The new pati...

Страница 58: ...inted 7 If no valid data is collected for a patient number is printed 8 If valid data is collected for a patient number for less than one minute only the last measurement is printed 9 If data is collected for a patient number for more than one minute the relative time since the first measurement is shown for that patient 10 Dashes indicate invalid or unavailable data for example the patient s fing...

Страница 59: ...itor functions operate as described previously in this manual NOTE Accessing Home Use Mode and Sleep Study Mode may only be done from the Clinician Mode It is not possible to access Home Use Mode from Sleep Study Mode or vice versa Home Use Mode While the monitor is in home use mode Alarm limits can be viewed but cannot be changed The alarm and alert tone is set to loud and cannot be changed Alarm...

Страница 60: ...ing the following commands a Turn off the monitor b Press and hold the ALARM VOL key then press the ON key c While holding the ALARM VOL key k flashes in the pulse rate display d When k stops flashing and lights steady in about 6 sec onds release the ALARM VOL key The monitor is now in home use mode 4 Verify the monitor is in the home use mode a Turn off the monitor b Turn on the monitor while obs...

Страница 61: ...boot with carrying case 3049 Adhesive Strips Adhesive Tape 1 1854 Home Use Instruction Book Quantity prescribed by doctor The doctor will prescribe the type and quantity of the sensors needed for home use QUANTITY CAT NO DESCRIPTION 3044 Sensor Reusable Finger 3178 Sensor Pediatric Finger 1 3311 Cable Oximetry 5 Feet 3444 Sensor Comfort Clip 1300 Sensor Disposable Adult 1301 Sensor Disposable Pedi...

Страница 62: ...or the hospital emer gency room 1 Emergency phone number for local paramedics or police 1 Phone number for equipment supplier in case the equipment fails Quantity prescribed by doctor Training the Home Use Caregiver 1 The home use caregiver must be trained in CPR 2 Make sure the monitor s alarm limits are properly set and that the monitor is in the home use mode 3 Inform the caregiver that the oxi...

Страница 63: ...ngth bar graph read ings Change the brightness of the display Change the pulse beep volume Turn off the alarm and alert tones for two minutes Turn on the alarm and alert tones Interpret the alarms View the alarm limits Interpret the SENSOR alert Interpret the BATT attention Turn off the monitor 5 Tell the caregiver how to respond In case of a patient emergency including what therapy to pro vide th...

Страница 64: ...key c While holding the PULSE VOL key k flashes in the pulse rate display d When k stops flashing and patient number pn n patient number lights steady in about 6 seconds release the key The monitor is now in clinician mode 2 Verify the monitor is in the clinician mode a Turn off the monitor b Turn on the monitor while observing the display the soft ware revision then the patient number should be d...

Страница 65: ...nd data from the patient is stored and may then be output to a PC Following the header the SpO2 and pulse rate trend block is out put in ASCII as follows Sleep Study SpO2 Pulse 97 67 98 69 98 71 98 71 Sleep Study SpO2 Pulse 97 67 98 69 98 71 98 71 Each pair of data points represents a four second block of data For each trend block the header Sleep Study appears preceding the trend block indicating...

Страница 66: ...ereby overwriting the oldest data the most recent data is always retained The serial communication setup is the same as in the clinician mode 9600 baud 1 start bit 8 data bits 1 stop bit no parity All other functions of the monitor work as in the clinician mode Setting Up the Monitor for Sleep Study Mode NOTE When entering or exiting Sleep Study mode previous memory will be lost 1 Put the monitor ...

Страница 67: ...the monitor b Press and hold the Alarm Silence key then press the ON key c While holding the Alarm Silence key SLP flashes in the pulse rate display d When SLP stops flashing and patient number pn n patient number lights steady in about 6 seconds release the key The monitor is now in clinician mode 2 Verify the monitor is in the clinician mode a Turn off the monitor b Turn on the monitor while obs...

Страница 68: ...rinter communication cable to the monitor 3 Set up the communication software to accept the following RS 232 data format Data Type ASCII Data Format 9600 baud 1 start bit 8 data bits 1 stop bit no par ity 4 Connect the standard RS232 null modem cable s DB 9 connector to the mating connector of the printer cable 5 Connect the standard RS232 null modem cable s DB connector to the mating connector on...

Страница 69: ... Set up the monitor cables and PC as previously described 2 Turn on the PC and load the communication software setting the serial communication defaults as previously described 3 Turn on the monitor 4 Dashes indicate invalid or unavailable data for example the patient s finger was removed from the SpO2 sensor ...

Страница 70: ...This page is intentionally left blank ...

Страница 71: ... and Accessories to verify the proper AC power supply for your application 2 Make sure the POWER indicator lights up on the charger If not see the Troubleshooting section 3 CHGING indicator lights while the monitor s battery is charging 4 Device is fully charged when the CHGING light turns off 5 The monitor s battery will fully charge in about four 4 hours NiCad or 6 hours NiMH A fully charged bat...

Страница 72: ...our installation site before using this monitor The monitor s AC rating is shown on the external power supply If the rating is not correct do not use the monitor contact the Smiths Medical PM Inc service department or your local dis tributor for help CAUTION Monitors not in use must be fully charged at least every five months otherwise the battery will no longer be able to accept a charge ...

Страница 73: ... Follow the instructions for attaching the sensor Disinfecting the 3044 reusable sensor Before attaching the sen sor to the patient Follow the instructions for attaching the 3044 reusable sensor Disinfecting the monitor and the AC power sup ply When necessary 1 Disconnect the moni tor from the AC power supply 2 Disconnect the AC power supply from the wall outlet and from the monitor 3 Wipe the sur...

Страница 74: ...e monitor s connector 2 Make sure the sensor is properly attached to the patient See Attach ing the Sensor to the Patient section for help 3 Make sure the adhesive tape used to hold the sensor is not wrapped too tightly Wrapping the tape too tightly may prevent the monitor from measuring the SpO2 and pulse rate 4 If the sensor alert is still on contact your doctor or the equipment dealer for help ...

Страница 75: ... Monitor SpO2 pulse rate or pulse bar graph not shown on monitor or Pulse rate value erratic intermittent or incor rect or SpO2 value erratic intermittent or incor rect 1 Patient may be moving Have patient remain as still as possible 2 Connectors from sensor to monitor may not be firmly seated Make sure all connec tors between sensor and monitor are firmly seated 3 Sensor may not be properly attac...

Страница 76: ...wall outlet and see if the lamp or radio turns on 4 AC power supply or monitor may be defec tive Contact authorized repair center for help CHGING does not light on the monitor 1 Monitor s battery may be charged No printout on optional printer 1 Printer power not connected or printer power switch is off Make sure the power is connected to the printer and the printer power switch is on 2 Printer int...

Страница 77: ... Book Oximeter each 3025 Sensor Oximetry Wrap Infant 3 15 Kg each 3026 Sensor Oximetry Wrap Neonate 3 Kg each 3043 Sensor Oximetry Universal Y each 3044 Sensor Oximetry Finger each 3049 Strips Adhesive 40 pkg 3078 Sensor Oximetry Ear each 3134 Tape Attachment Neonatal 50 pkg 3135 Tape Attachment Infant 50 pkg 3136 Tape Attachment Neonatal 100 pkg 3137 Tape Attachment Infant 100 pkg 3138 Posey Wrap...

Страница 78: ...n Manual Ordering Information For ordering information contact your local distributor or the Smiths Medical PM Inc customer service department Smiths Medical PM Inc N7 W22025 Johnson Drive Waukesha WI 53186 USA Phone 262 542 3100 Fax 262 542 3325 Email Address info smiths bci com ...

Страница 79: ...graph SENSOR Sensor alert indicator BATT Low battery indicator Silenced Alarm and alert tone silenced indicator Keys Nine control keys provided Brightness Adjustable brightness of SpO2 pulse rate and bar graph dis plays Indicators POWER Indicates oximeter connected to AC power CHGING Indicates oximeter battery is charging SpO2 Range 0 100 Accuracy 2 at 70 100 3 at 50 69 Alarm Limits High 100 50 an...

Страница 80: ...not user replaceable Charge Time Nicad Fully charges in about 4 hours NiMH Fully charges in about 6 hours Use Time Approximate continuous use 12 hours Nicad 24 hours NiMH Indicators BATT indicator lights when about 30 minutes of battery use remains AC Charger Wall Mount Style Input of 105 125 V 60Hz Input of 100V 50Hz Optional Table Top Style Input of 230 V 50 60Hz Dimensions Width 3 3 inches 84 m...

Страница 81: ......

Страница 82: ...nc Patient Monitoring and Ventilation m Authorized Representative as defined by the Medical Device Directive Smiths Medical International Ltd Colonial Way Watford Herts UK WD24 4LG Phone 44 1923 246434 Fax 44 1923 240273 ...