Instruction Manual ColorPlus Ex
10210E/5
21
The maximum distance (y) between SITRA and photometer depends on the
photometer type and the cable cross-section used:
Cable cross-
section [mm2]
ColorPlus Ex UV
Max. length y [m]
ColorPlus Ex VIS
Max. length y [m]
0.14
16
48
0.25
30
87
0.34
41
119
0.5
59
171
0.75
87
256
standard cable
1
114
333
1.5
162
476
The maximum distance (z) between SITRA and control unit depends on the
cable cross-section used:
Cable cross-section [mm2] ColorPlus Ex UV + VIS
Max. length z [m]
0.14
471
0.25
844
0.34
1161
0.5
1200
0.75
1200
standard cable
1
1200
1.5
1200
3.6
Installation of the SIREL Ex control unit
Please consult manufacturer documentation 11044DEF when installing the
SIREL Ex.
A detailed dimension drawing of the SIREL Ex is in Section 11.
Refer to the table in Section 3.5 for distance dependency on cable cross-
section.