Instruction Manual ColorPlus Ex
10210E/5
57
7
Taking Out of Service/storage
Do not open the photometer before the power supply to the control unit has
been disconnected and then a waiting time of ten minutes has passed to allow
the light source to cool down sufficiently (EXPLOSION HAZARD)!
(Cooling
down of the device to a safe temperature level and elimination of residual
charges.)
Action
1.
Switch off the power supply to the control unit and disconnect all of
the electrical connections to the control unit.
2.
Dismount the control unit.
3.
Thoroughly clean all surfaces that have come into contact with the
medium. Now no toxic, corrosive or loose deposits should remain
inside.
4.
Make sure that all covers and hoods are closed and all closures on the
photometer and control unit are secured.
No special conditions are required for storage of the equipment, but be sure to
observe the following:
The photometer and control unit contain electronic components. The
storage must satisfy the normal conditions required for such components.
In particular, the storage temperature should remain within the –20 ..
50°C range.
During storage, the photometer, control unit and accessories must be
protected against weathering, condensing moisture and aggressive gases.
Taking out of service
Storage of the
equipment
Содержание ColorPlus Ex
Страница 2: ...Sigrist Gruppe Subject to change without notice 09 19...
Страница 6: ...Instruction Manual ColorPlus Ex iv 10210E 5...
Страница 63: ...Instruction Manual ColorPlus Ex 10210E 5 61 11 Appendix...
Страница 64: ...Instruction Manual ColorPlus Ex 62 10210E 5...
Страница 66: ...Instruction Manual ColorPlus Ex 64 10210E 5...