Instruction Manual ColorPlus Ex
38
10210E/5
5.2
100% adjustment of the photometer
Recalibration of the photometer can change the readings slightly from the
previous ones, because the instrument is reset to a reference standard (e.g.
distilled water)!
Before the 100% adjustment is started, the photometer must have reached the
operating temperature (UV > 2h/LED > 10 min)!
Action
Display (example)
Remarks
1.
Fill flow cell with zero
medium
1
e.g. with
distilled water or air
K1 00.70 E
K2 90.50 %T
Make sure there are
no air bubbles in the
flow cell during the
zero adjustment.
2.
Activate service mode.
* 100% ADJUST. *
* *
Section 4.3
3.
2 x
> 100% Adjust. <
No
4.
Activate editing mode
with
.
100% Adjust.
> No <
5.
1 x
or
100% Adjust.
> Yes <
6.
Confirm selection with
.
Adjustment
running ...
The process takes
about 30 seconds...
7.
M100.0 100.2 99.8
K 0.0 0.1 -0.0
Instrument in service
mode.
M = reading for
transmission (must be
around 100%)
K = compensation
(must be around 0)
8.
+
(together)
K1 00.10 E/m
K2 100.1 %T
Instrument in normal
operation.
If the 100% adjustment could not be carried out, the message
100%
adjustment outside tolerance
appears in the display. If that occurs,
check the following points:
Ensure the device is properly mounted
The measuring cell may be too dirty
Ensure the correct zero medium is used
1
In the UV range (254nm, 313nm) distilled water is used as zero medium; in
the visible range (365..700nm) either clear, colorless water or any colorless
sample else is normally adequate enough.
Содержание ColorPlus Ex
Страница 2: ...Sigrist Gruppe Subject to change without notice 09 19...
Страница 6: ...Instruction Manual ColorPlus Ex iv 10210E 5...
Страница 63: ...Instruction Manual ColorPlus Ex 10210E 5 61 11 Appendix...
Страница 64: ...Instruction Manual ColorPlus Ex 62 10210E 5...
Страница 66: ...Instruction Manual ColorPlus Ex 64 10210E 5...