Instruction Manual ColorPlus Ex
10210E/5
25
Make up the electrical connections in the following order:
Terminals
Used for
Remarks
1.
20
Connection to photometer green
21
brown
22
white
23
yellow
2.
4 - 5 - 6
Relay output 1
The relay outputs are
freely configurable
(
Reference Handbook).
7 - 8 - 9
Relay output 2
18 - 19
Reading output for
1
st
meas. angle
(90° reading in the case of
dual-angle instrument)
0/4 .. 20 mA, max.
burden 600
If unused, these
terminals must be
shorted with a jumper!
32 - 33
Reading output for
2
nd
meas. angle (25°
reading in the case of dual-
angle instrument)
3.
10 .. 17
Control signals for
1
st
meas. angle
Reference Handbook
26 .. 31
Control signals for
2
nd
meas. angle
Provided on dual-angle
instrument only.
16a
Control signal for sensor
check
Reference Handbook
24 - 25
Connection to bus coupler
Reference Handbook
4.
1 - 2 - 3
Mains voltage
85 .. 264 V; 47 .. 63 Hz
or 24 VDC; 25 W
Connection