Instruction Manual ColorPlus Ex
10210E/5
41
5.4
Replacement of the desiccant in the sensor head
Action
Remarks
1.
Do not open the photometer before the
power supply to the control unit has been
disconnected and then a waiting time of ten
minutes has passed to allow the light source
to cool down sufficiently (EXPLOSION
HAZARD)!
(
Cooling down of instrument to a safe
temperature level and fading of the residual
charges.)
Disconnect the power supply of the
photometer to the control unit.
2.
Remove the six screws (A) around the
periphery of the sensor head.
3.
Withdraw the sensor head from the adapter
ring.
4.
Loosen the three screws (A) and remove the
leakage protector.
5.
Remove the old desiccant.
Place the new desiccant (A) in the leakage
protector and then cover it with the dust
protector (B).