Instruction Manual ColorPlus Ex
30
10210E/5
4.2
Normal operation (operating mode)
When the instrument is switched on, ColorPlus Ex first checks which control
units are connected to it. Then it switches automatically to normal operation.
If several SIRELs are connected to the photometer, all of the control units must
be switched on for operation!
The following displays appear in normal operation:
Display (Examples)
Signifies
K1 92,80 E/m
The reading of channel 1 is displayed. There is no
second channel.
K1 90,60 E/m
K2 95,20 E/m
The readings of channels 1 and 2 readings are
displayed. Use
to switch to the display for
channel 3, if any. The math channel is not active.
M1 90,60 E/m
or
M2 95,20 E/m
Math channel 1 (or math channel 2) is active and is
displayed. Use
to switch to the display of the
two channels 1 and 2. Press
again to show the
display for channel 3, if any.
M1 90,60 E/m
M2 95,20 E/m
Both math channels are active. Use
to switch to
the display of the two channels 1 and 2. Press
again to show the display for channel 3, if any.
K1 0.000-3.000
Display of the actual measuring ranges (press
to
switch).
Besides the display shown during normal operation, it is also possible to show
the following displays:
The display...
signifies ...
so you should...
K1 ***** E/m
K2 ***** E/m
K3 ***** E/m
M1 ***** E/m
M2 ***** E/m
...that the reading is
outside the momentary
measuring range.
...make sure that the
absorption lies within the
set range (max. 3E). If
necessary, alter the path
length.
... ignore the display if
the process happens to
be in an irregular phase.
**** Fault ****
...
...that a malfunction has
occurred.
...try to pinpoint the
cause of the malfunction
(
Section 6).
**** Warning ***
...
...that a warning
situation has occurred.
...try to determine the
cause of the warning
(
Section 6).
Table 2: Display examples and what they signify.
Displays in normal
operation
Special displays
Содержание ColorPlus Ex
Страница 2: ...Sigrist Gruppe Subject to change without notice 09 19...
Страница 6: ...Instruction Manual ColorPlus Ex iv 10210E 5...
Страница 63: ...Instruction Manual ColorPlus Ex 10210E 5 61 11 Appendix...
Страница 64: ...Instruction Manual ColorPlus Ex 62 10210E 5...
Страница 66: ...Instruction Manual ColorPlus Ex 64 10210E 5...