Instruction Manual ColorPlus Ex
54
10210E/5
6.3
Warnings
Temperature, leakage and degree of contamination are monitored continuously
in the ColorPlus Ex. If this monitoring shows that a limit has been exceeded, a
warning is emitted and displayed on the control unit.
These warnings can be detected by ColorPlus Ex:
Message
Signifies
Possible causes
Warning
Temperature
The temperature in the
transmitter has exceeded
65°C
Excessively high medium
or ambient temperature
and a defective cooling
system or none at all.
Warning
Sealing
The relative humidity in
the transmitter (and
possibly also in the
receiver) has risen above
the preset limit
The desiccant is
saturated (
Sections
Seals are no longer
sealing properly
Warning
Soiling
Contamination of the
flow cell has exceeded
the preset limit (only
present in the version for
water).
Badly contaminated
medium
Cleaning interval too
long
Incorrect installation
position (
Section 3)
Warning
Neg.Ext.
The measuring value has
exceeded the negative
limit (Neg.Ext. LI)
The device is dirty
(
Table 7: Warnings