Instruction Manual ColorPlus Ex
10210E/5
17
3.2.2
Installation with shortened path length
The ColorPlus Ex is delivered for installation in an inline housing with a
shortened path length (7) on both sides. The path length shorteners (OPL bits)
with flow cell window (3), matching leakage protector (8) and adapter ring (2)
are installed at the factory to meet the customer’s requirements.
Action
Remarks
1.
Insert the completely assembled transmitter /
receiver (including flow cell window (3),
adapter ring (2), leakage protector (8) and
screwed-on OPL bit) into the receptacle of the
inline housing.
The groove (A) on transmitter and receiver
must be installed pointing upward!
2.
Fasten the transmitter / receiver with the
hinged ring (4) and tighten nut (A).
Mounting of
transmitter and
receiver