Instruction Manual ColorPlus Ex
48
10210E/5
Action
Remarks
7.
Take the thrust ring and the flow cell window
with seal out of the OPL bit. Remove the seal
from the flow cell window and clean the
window with a cleaning paper or rag, possibly
using warm soapy water. Take care not to
dirty the window again during reinstallation. If
necessary, replace the window.
8.
If necessary, replace the seal (B). Place the
seal into the window’s groove. Insert the
window into the OPL bit with the seal facing
downward.
9.
Place the thrust ring into the OPL bit with the
ring’s groove facing downward.
10.
Screw in the threaded ring and tighten it with
the OPL bit wrench.
11.
If necessary, place a new seal (arrow) into
the groove on the OPL bit.
12.
If necessary, replace the leakage protector
seal (A). Remove the old seal and place the
new seal into the groove on the leakage
protector.
If necessary, replace the desiccant and
the seal (
Section 5.4).
13.
Slip the adapter ring (with the screwed-on
OPL bit) on the transmitter / receiver and
fasten it with the six screws.