Troubleshooting
6.1
Pinpointing the cause of a malfunction
To narrow down the cause of a malfunction, proceed step by step according
to the following table. If the action taken fails to produce the desired result,
consult Customer Service (
Section 6.5).
Perceived fault
Action
No display
Make sure the control unit is properly
connected to the supply voltage
(
Section 3.7)
Check the fuses in the control unit
(
Reference Handbook)
Fault message or warning
in the display
Analyze the fault message / warning
(
The reading seems
suspect
Check whether the photometer is correctly
mounted (
Section 3.1)
Make sure the servicing chores have been
carried out properly in accordance with the
Service Schedule (
Section 5.1).
Carry out a 100% adjustment on the
photometer (
Section 5.2)
Replace the desiccant in the sensor head
(
Section 5.4).
Carry out a functional check with the checking
glass (
Section 5.3).
No measured value output
signal (0 mA)
-
This status signals that a fault has occurred in
the photometer.
-
Analyze the fault displayed fault message
(
Section 6.2).
Table 5: Pinpointing a malfunction
6.2
Fault messages
In the ColorPlus Ex, readings are measured, inputs monitored and outputs
updated continuously. If the ColorPlus Ex detects a fault during this process, a
fault message will be emitted and displayed on the control unit.
Содержание ColorPlus Ex
Страница 2: ...Sigrist Gruppe Subject to change without notice 09 19...
Страница 6: ...Instruction Manual ColorPlus Ex iv 10210E 5...
Страница 63: ...Instruction Manual ColorPlus Ex 10210E 5 61 11 Appendix...
Страница 64: ...Instruction Manual ColorPlus Ex 62 10210E 5...
Страница 66: ...Instruction Manual ColorPlus Ex 64 10210E 5...