Instruction Manual ColorPlus Ex
10210E/5
31
4.3
Service mode
The photometer can be configured in the service mode. Measuring is
interrupted. The following menu control appears on the display:
Action
Display (example)
Remarks
1.
Access code
> 000000 <
If no personal access
code has been set,
confirm with
. Then
continue with step 3.
2.
Enter code:
/
Change number
/
Change place
Access code
> ...... <
Now enter your own
access code, or the
factory default
000000.
3.
* 100% ADJUST. *
* *
Instrument is in the
service mode.
Depending on the configuration, the reading output either switches to 0/4 mA
or remains frozen at the last reading (
Reference Handbook).
4.4
Set your language
The control unit’s display can be set to English, German, French or Italian.
Action
Display (example)
Remarks
1.
Switch to service
mode.
* 100% ABGL. *
* *
Section 4.3
2.
3 x
to
* KONFIGURIEREN*
* *
3.
> Sprache <
Deutsch
4.
Activate editing mode
with
.
Sprache
> Deutsch <
5.
Select language:
/
Sprache
> ... <
6.
Confirm selection with
.
> Language <
...
7.
+
(together)
K1 68.2 %T
K2 85.8 %T
Instrument in normal
operation.
Applies in service
mode: