Instruction Manual ColorPlus Ex
10210E/5
19
3.5
Installation of the control units SIREL SMD
Do not install or operate the control unit or any supplementary components in
areas at risk of explosion (CAN BE FATAL)!
The control units can be mounted right on the wall, on a mounting frame or on
a stand.
The standard cable supplied is 5 m long. This ensures that the control unit will
be positioned close to the photometer.
Pos.
Name
1
Photometer
2
SIREL
control unit
3
Non Ex-zone
4
Partition
5
Ex-zone
Figure 11: Control unit position with standard cable
For distances longer than 5 m it is necessary to insert an additional Ex-tested
terminal box between photometer and control unit. This permits quick
connection and disconnection of the photometer and control unit.
Pos. Name
1
Photometer
2
Ex-tested
terminal box
3
SIREL control
unit
4
Non Ex-zone
5
Partition
6
Ex-zone
Figure 12: Arrangement of control units located at distances longer than 5 m
Mounting SIREL
Distance up to 5 m
Distances longer
than 5 m – variant 1
Содержание ColorPlus Ex