Instruction Manual ColorPlus Ex
40
10210E/5
Action
Remarks
9.
Press key
.
Depending on the configuration, between one
and three transmission values will appear in
the top line.
10.
Take the reading(s) and note it down.
11.
Compare the measured value with the value
noted down from inside the enclosure.
If the variance is more than ±0.1 %T, repeat
steps 1
−
11. If the measured value varies
too much again, contact your service office.
If the measured value agrees with the value
stated inside the enclosure, proceed with
step 12
.
12.
Do not open the photometer before the
power supply to the control unit has been
disconnected and then a waiting time of ten
minutes has passed to allow the light source
to cool down sufficiently (EXPLOSION
HAZARD)!
Disconnect the power supply of the
photometer to the control unit.
13.
Loosen nut (A) enough to be able to swing
the hinged ring away easily. Take off the
entire clamping ring and then remove the
receiver enclosure.
14.
Loosen knurled nut (A) and turn the checking
glass back to the normal position. Tighten
knurled nut (A).
The checking glass must be in contact
with stop (B)!
15.
Reclose the instrument in the reverse order.