Installation/start-up
3.1
Photometer installation for the bypass version
The photometer must be installed in the horizontal position. To make it
possible to vent the flow cell properly, make sure the medium outlet is on top.
Action
1.
Fasten the photometer horizontally to a
wall with four screws on fastening plate
(A). The transmitter (B) must be on the
right, the receiver (C) on the left.
2.
Now connect the sample inlet and
outlet.
D: Medium inlet
E: Medium outlet
Mounting of the
bypass version