Instruction Manual ColorPlus Ex
32
10210E/5
4.5
Set the measuring range
Action
Display (example)
Remarks
1.
Activate service mode.
* 100% ADJUST. *
* *
Section 4.3
2.
* MEAS.RANGE *
* LIMITS *
3
.
> Meas.range <
X...
Set the range for
output signals 1 + 2.
4
.
Activate editing mode
with
.
Meas.range
> X... <
5.
Select measuring range
with
/
.
Meas.range
> 1.. <
e.g. range 1
6.
Confirm selection with
.
> Meas.range <
1..
7.
+
(together)
K1 68.2 %T
K2 85.8 %T
Instrument in normal
operation.
Measuring range No.
Measuring span in E
User-specific
MR 1 (Standard setting)
0 .. 3
MR 2
0 .. 2
MR 3
0 .. 1
MR 4
0 .. 0.5
MR 5
0 .. 0.2
MR 6
0 .. 0.1
MR 7
0 .. 0.05
MR 8
0 .. 0
MR inputs
Remotely controlled range switching
Reference Handbook
Automatic
0 .. 3
Table 3: Measuring ranges
The measuring ranges are set at the factory to suit the user’s requirements.
Depending on the instrument version, certain measuring ranges can be masked
out (
Configuration sheet).
It is also possible to modify the measuring ranges individually (
Reference
Handbook). Changes can be entered in Table 3: Measuring ranges in the
column “User-specific” as well as noted on the parameter list.
Measuring ranges set
at the factory
Содержание ColorPlus Ex
Страница 2: ...Sigrist Gruppe Subject to change without notice 09 19...
Страница 6: ...Instruction Manual ColorPlus Ex iv 10210E 5...
Страница 63: ...Instruction Manual ColorPlus Ex 10210E 5 61 11 Appendix...
Страница 64: ...Instruction Manual ColorPlus Ex 62 10210E 5...
Страница 66: ...Instruction Manual ColorPlus Ex 64 10210E 5...