Instruction Manual ColorPlus Ex
46
10210E/5
Action
Remarks
7.
Now you can remove the threaded ring,
thrust ring, seal and flow cell window.
8.
Clean the flow cell window with a suitable
cleaning agent. Take care not to dirty it again
during reinstallation. If necessary, replace the
flow cell window and seal.
9.
If necessary, replace the leakage protector
seal (A). Remove the old seal and place the
new seal into the groove on the leakage
protector.
If necessary, replace the desiccant and
the seal (
Section 5.6).
10.
Reassemble the photometer in reverse order
and restart it.
11.
Carry out a 100% adjustment as directed in
Section 5.2.
5.8
Cleaning/replacement of the flow cell windows
on the Varivent
®
flow cell
Action
Remarks
1.
Interrupt the sample flow through the in-line
flow cell and let it run empty. Switch off the
photometer’s power supply.
2.
Remove first nut (A) and then the hinged
rings on the transmitter and receiver sides.
Get a firm grip on the transmitter /
receiver while doing this!