Instruction Manual AquaScat 2 P
Servicing
52
11838E/2
8.6.
Check and clean optional accessory parts
Damage to components due to aggressive cleaning agents.
Do not use aggressive cleaning agents when cleaning optional accessory parts. Commercially
available dishwashing liquid can be used if needed.
When checking for cleanliness of the optional accessory parts, the following components
must be examined for particle residues:
Flow meter with regulator valve
Hoses
If particle residues are found, rinse the concerned components with clean water. Use com-
mercially available dishwashing liquid if needed.
8.7.
Clean the closed measuring cell
Damage to instrument due to aggressive cleaning agents.
Do not use aggressive cleaning agents or solvents to remove lime deposits on the instru-
ment.
The fold-out Figure 20 is available in the Appendix to assist you when performing servicing
duty.
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Interrupt the sample supply to the pho-
tometer.
2
Remove sample connections and wait
until the measuring cell empties.
3.
Remove the photometer from the meas-
uring position and fasten onto the dock-
ing station.
Section 8.2
CAUTION!
CAUTION!