Instruction Manual AquaScat 2 P
Servicing
46
11838E/2
8.3.
Remove the measuring cell unit
WORKSTEP
PROCEDURE
1. Fasten photometer to the
docking station.
Section 8.2
2. Loosen the five fastening
screws of the measuring
cell unit (circles).
3. Loosen mounting clip and
remove optics unit
3.1: Use a little force to push the red fuse in the direction
of the arrow (picture 1) and at the same time lift the
mounting clip (picture 2).
Picture 1
Picture 2
3.2: Press mounting clip in the direction of the arrow
over the lock plate of the optics unit (picture 3) and then
open (picture 4).
Picture 3
Picture 4
3.3: Remove optics unit from the measuring cell unit.
4. Remove measuring cell unit
and place on a safe surface.