Servicing
Instruction Manual AquaScat 2 P
11838E/2
45
WORKSTEP
PROCEDURE
1. Interrupt sample supply to
the photometer and wait
until the measuring cell
empties.
2. Fasten photometer to the
docking station.
2.1: Remove the photometer (Figure 18, 1) from the
measuring position and place on the docking station Fig-
ure 18, 2).
Pay attention to the milled area (dark colored area).
2.2: Fasten the photometer (Figure 18, 1) with knurled
screw (Figure 18, 3) to the docking station.