Instruction Manual AquaScat 2 P
Troubleshooting
56
11838E/2
WORKSTEP
PROCEDURE
3.
Remove the old fine-wire fuse (circle) from
the base print (AQBasi) and replace with a
new one (type T2A).
4.
Carefully mount front cover and fasten
with the five screws.
Damage to the threaded inserts in
the housing due to excessive tighte n-
ing of the screws when fastening the
front cover:
Use a hex key without a T-handle to tight-
en the screws of the front cover finger-
tight (approx. tightening torque 1 Nm).
7 mm hex key
5.
The instrument can now be operated
again.
9.3.
Warning and error messages
9.3.1.
Warning messages and effect on operation
WARNING:
If a warning occurs during operation, it has the fol-
lowing effects:
Warnings indicate an unusual state.
The system continues to operate and delivers
correct measuring values. The cause of the
warning message should be remedied at the
next possible opportunity.
If an output for warnings is programmed, it
is activated.
When the cause of the warning has been
remedied, it is automatically deleted.
When the
Warning
message
occurs,
the color of the status display changes
to
orange
and the warning text de-
scribes what the warning is about.
(Example:
HUMIDITY WARNING
)