Instruction Manual AquaScat 2 P
Operation
36
11838E/2
7.8.
Setting the current outputs
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Curr. outputs
button.
4.
Select current output 1 or 2.
5.
Select
C1 Turb
source.
For
Inactive
the current output is
switched off.
6.
Select range.
MR1 .. MR8 (see table below for range
definition)
In 1, In 2, Auto 1, Auto 2.
Reference Manual
7.
Press the
Meas
button.
Instrument again in measuring opera-
tion.
MEASURING
RANGE NO.
MEASURING RANGE
DEFAULT
MEASURING RANGE
(CUSTOMER SPECIFIC)
1
0 .. 100
2
0 .. 50.0
3
0 .. 30.0
4
0 .. 10.0
5
0 .. 3.00
6
0 .. 1.00
7
0 .. 0.30
8
0 .. 0.10
If other measuring ranges are needed, you can re-program the table above as required.
Reference Manual