Troubleshooting
Instruction Manual AquaScat 2 P
11838E/2
55
9.
Troubleshooting
9.1.
Pinpointing the cause of a malfunction
DETECTABLE
MALFUNCTION
MEASURES
No reading.
Check whether the service voltage is connected.
Sec-
tion 5
Check whether the fine-wire fuse in the instrument is
OK.
Section 9.2
Fault report in the display.
Analyze the error message.
Section 9.3
The measuring value ap-
pears wrong.
Make sure there are no air bubbles in the sample.
Counterpressure may help.
Section 4.4.1
Ensure that the sample to be measured corresponds to
the operating conditions.
Section 2.4
Perform adjustment.
Section 8.5
Check whether the photometer is correctly mounted.
Section 4
Ensure that the maintenance work has been performed
according to the servicing schedule.
Section 8.1
Clean measuring cell.
Section 8.7
If the listed measures do not result in the desired results, please consult with customer se r-
vice.
Section 10
9.2.
Replace fine-wire fuse
WORKSTEP
PROCEDURE
1.
Interrupt the service voltage to the pho-
tometer.
Section 5
2.
Loosen the five screws and remove the
front cover.