Servicing
Instruction Manual AquaScat 2 P
11838E/2
47
8.4.
Replace desiccant
Figure 19: View of the measuring cell element
Desiccant (4x)
Measuring cell
Support
Measuring cell plate
WORKSTEP
PROCEDURE
1.
Interrupt the sample supply to
the photometer.
2.
Remove sample connections and
wait until the measuring cell
empties.
3.
Remove measuring cell unit and
place on a safe surface.
Section 8.3
4.
Remove the old desiccant
Figure 19, 1) on both sides and
replace with new desiccant.
Two desiccants on each
side.