WORKSTEP
PROCEDURE
1. Pre-mounting the mounting
bracket.
Screw on the mounting bracket (Figure 3, 2) horizontally
at the intended position.
2. Fasten the photometer onto
the pre-mounted mounting
bracket.
2.1: Position the photometer on the pre-mounted
mounting bracket (Figure 3, 2) and pay attention to the
two positioning pins (arrows, see picture).
2.2: Screw the photometer onto the mounting bracket
(Figure 3, 2).
3. Align the photometer.
3.1: Loosen the fixing nut (Figure 3, 4).
3.2: Adjust the support (Figure 3, 3) so that it contacts
the wall and relieves the load stress on the instrument
(Figure 3, 1).
3.3: Tighten the fixing nut (Figure 3, 4).