Operation
Instruction Manual AquaScat 2 P
11838E/2
39
7.10.
Configuring the flow meter
The optional flow meter is connected to terminals 22, 23 (In1). Proceed as follows to activate
monitoring:
MANIPULATION
SUPPLEMENTARY INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
Press the
Inp./outputs
button.
4.
Press the
Inputs
button.
5.
Select
Input 1
.
6.
Activate input functions
External
and
In-
vert
.
7.
Press the
OK
button.
8.
Press the
General
button.
9.
Under
Name ext.in
you can enter the text
of the error message with a maximum of
seven characters (e.g. FLOWRAT).
Prio.ext.in
should be on
Fault
.
10. Press the
Meas
button.
Instrument again in measuring opera-
tion.