Operation
Instruction Manual AquaScat 2 P
11838E/2
31
7.5.
Switching to service operation
The photometer is configured in service operation. The measuring procedure is interrupted
and the main menus appear on the display.
Proceed as follows for service operation:
MANIPULATION
SUPPLEMENTARY
INFORMATION
1.
Press the
Menu
button.
2.
Set the access code and confirm with
OK
.
Factory setting is
0
.
3.
The main menus appear.
Now you are in service operation.
The following applies in service operation:
* The measuring values remain on the last values on the digital interfaces.
* Depending on the configuration, the current outputs go to 0/4 mA or remain on
the last measuring value.
The limits are deactivated.
If an output for service is programmed, it is activated.
Error messages are suppressed.
* This applies when the
Curr.outputs\General\For service
parameter is set to
Measure
.
For measuring operation press the
Meas
button. When switching from service operation to
measuring operation, an hourglass appears in the information bar for about 20 seconds. The
measuring values are frozen during this time.
Definition of service
operation