Servicing
Instruction Manual AquaScat 2 P
11838E/2
49
8.5.
Manual adjustment
The manual adjustment should be carried out after the measuring cell is cleaned.
The fold-out Figure 20 is available in the Appendix to assist you when performing servicing
duty.
WORKSTEP
PROCEDURE
1. Interrupt the sample supply to
the photometer.
2
Remove sample connections
and wait until the measuring
cell empties.
3. Insert the checking unit into
the measuring cell.
3.1: Loosen the screw cap (Figure 20, 3) and remove
the measuring cell plate.