Maintenance
6-8
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
Error
1
)
Diagnosis
Remedy
F
Flashing ”F” in the top
right hand corner of the
display
General incorrect measurements,
process errors or incorrect
parameterization (see also
section 5.3.2)
Menu
2.1
Call device status
F
Device status (menu
2.1
)
”Measurement module
failure”
Measuring range overflow
Reduce flow quantity or increase full scale
value in menu
3.1.2
F
Device status (menu
2.1
)
”Measurement module
Measuring electronics defective
Measurement with reduced accuracy still
possible;
replace complete electronics
Defective immersion check
Deactivate immersion check (menu
2.3
)
Magnetic field current outside the
calibrated range
Recalibrate unit
Fault in measuring function
(incorrect measuring frequency)
Turn
off
mains
synchronization
(menu
3.1.7.2
).
F
Device status (menu
2.1
)
”Calibration failure
Self-test made with the result:
measuring tolerances too high in
Electronics must be recalibrated, send in
device or complete electronics or repair
”Calibration failure
measuring module”
measuring tolerances too high in
relation to factory calibration
device or complete electronics or repair
F
Device status (menu
2.1
)
”Simulation is running”
Device not in ”Measure” status,
Simulation is switched on
End simulation (exit the menu ”Simulation”
2.5.2
,
2.5.3
or
2.5.
4 or select ”End” in menu
2.5.1.
Alternatively: Restart the device
F
Device status (menu
2.1
)
”Calibration is running”
Zero calibration (menu
6.6.2
)
started and active
Device returns to measuring mode
automatically at the end of calibration
F
Device status (menu
2.1
)
”Tube empty”
(only possible with
Measuring tube empty or only
partially filled
Fill measuring tube with medium, ensure it is
completely full
(only possible with
immersion check
(menu
2.3
) activated)
High percentage of gas
Clear gas occlusions or use gas trap
F
Device status (menu
2.1
)
”Fl
t
Electrode DC voltage outside
itt d
Clean sensor electrodes
(
)
”Flow measurement
uncertain”
g
permitted range
uncertain
F
Device status (menu
2.1
)
Current flow too high
Reduce flow
(
)
”Measuring range
overflow”
Full scale value set too low
Increase full scale value in menu
3.1.2
or reduce flow rate
Table 6-2
Notes on troubleshooting
6.4
Cleaning
Use only
mild
cleaning agent to clean the transmitter. “Aggressive”
agents such as hydrochloric acid (even diluted), alcohol or abrasive
media may not be used.
Содержание SITRANS F M Intermag 2
Страница 2: ......
Страница 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Страница 9: ...1 1 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 Information for the User 1 ...
Страница 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...