Installation
3-9
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
S
Note that you may only use cables with at least the same degree of
protection as the sensor to install the sensor.
S
The line length from the cable gland to the terminals must be kept
as short as possible, line loops in the terminal box must be avoided.
S
To guarantee the IP 67 degree of protection, please use cables with
the following external diameters:
Auxiliary power and signal cable (large terminal box):
permissible external diameter
Cable gland
Standard
devices
7ME5033-xxAxx
Terminal box
explosion-proof
7ME5033-xxCxx
Terminal box in
increased safety
7ME5033-xxDxx
7ME5033-xxGxx
M20 x 1,5
6 - 12 mm
8 - 11,7 mm
5 - 9 mm
1/2” - 14 NPT
6 - 12 mm
6 - 12
mm
6 - 12 mm
Magnetic field current and electrode cable (small terminal box, only
remote model):
-
Cable glands M16 x 1.5
6 to 10 mm
-
Cable glands 1/2” NPT
5 to 9 mm
S
Connect the cables as follows:
-
Release the catch on the lid of the terminal box using a 3 mm
Allen key (only devices with a 100 to 230 V AC power supply and
devices with explosion protection type EEx de).
-
Unscrew the lid of the terminal box.
-
Push the supply cable and signal cable through the cable glands
up to the terminal strip. Lay the cable in a loop before the cable
glands so that moisture does not get inside the terminal box
(Figure 3-9).
-
Connect the lines according to Figure 3-11 through Figure 3-13.
Fit end ferrules to fine-wire lines.
-
Tighten the cable gland and check strain relief.
-
Screw the lid to the housing and tighten it. You must not use any
tools. The sealing ring must be clean and undamaged.
-
Mount the lid catch.
-
For transmitters in remote design, also connect the housing to
the local potential equalization to which the appropriate sensor
must also be connected.
Содержание SITRANS F M Intermag 2
Страница 2: ......
Страница 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Страница 9: ...1 1 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 Information for the User 1 ...
Страница 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...