Operation
5-41
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
Customer Code
You can change the personal code number in this menu item.
menu item
6.2
The code is factory set to 0. The menu item is only
accessible after entering the personal code even when programming
has already been enabled.
Attention!
The programming is always enabled when the code = 0.
Service code
The calibration data of the device are protected by a special code
menu item
6.3
(factory code). The appropriate menu items are only available after
entering this factory code.
Reset
Here you can perform a “warm start” of the controller,
menu item
6.4
This retains all saved units but the volume totalizer is reset to zero.
Control values
These display values are for trained service personnel,
menu item
6.5
since they are device-internal data mainly for error diagnosis.
These include:
S
Magnet voltage U
m
, menu item
6.5.1
S
Signal voltage U
sig
, menu item
6.5.2
S
Electrode voltage U
el1
, menu item
6.5.3
S
Electrode voltage U
el2
, menu item
6.5.4
S
Current flow
in % of the set full scale value, menu item
6.5.5
S
Measuring frequency
, menu item
6.5.6
S
Self test
, menu item
6.5.7
S
Service information
, menu item
6.5.8
S
Digital input
, menu item
6.5.9
Zero Trim
This function is divided into:
menu item
6.6
S
Zero flow correction
, menu item
6.6.1
This menu item serves to adapt the local hydraulic conditions. The
zero point has a speed value of
0 ... 1 m/s added. This setting is
undone by entering “0”.
S
Start correction
, menu item
6.6.2
This automatically determines the hydraulic zero point. This value is
visible in menu item
6.6.1
and can be adapted there if necessary or
undone by entering “0”.
Содержание SITRANS F M Intermag 2
Страница 2: ......
Страница 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Страница 9: ...1 1 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 Information for the User 1 ...
Страница 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Страница 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...