9-2-4 Sampling Mode Settings
Parameter: [SAMPLING MODE]
Setting Parameter
Description
Settings Range
SAMPLING MODE
This selects the measurement method.
x
CONTINUOUS:
Continuous
measurements
x
TIMED:
Timed measurements
CONTINUOUS or
TIMED
(Default setting:
CONTINUOUS)
START AT
This is the starting time for timed
measurements.
Date in yyyy/mm/dd
format
HH:mm:ss
DURATION
This specifies how long to continue timed
measurements.
Days, hours, and
minutes
(Max. 99 days, 23 hours,
and 59 minutes)
[SAMPLING MODE] Screen Operations
Use the up/down arrow keys to select [CONTINUOUS] or [TIMED] and select [SAVE] to
confirm (by pressing [1]).
[SAMPLING MODE] Screen
Note: If [TIMED] is specified, the [START AT/DURATION] screen is displayed.
[START AT/DURATION] Screen Operations
Use the left/right arrow keys to move the cursor and the up/down arrow keys to change
setting values. After configuring settings, select [SAVE] (by pressing [1]).
[START
AT/DURATION]
Screen
Measurement mode
Measurement start time
Measurement duration