38
(3) Select [START] (by pressing [1]).
→
A confirmation screen is displayed.
Operation Screen (Standby mode)
Confirmation Screen
(4) Select [START] (by pressing [2]).
→
The pump starts operating (the [MEASURE] lamp illuminates) and the measurement starts.
Confirmation Screen
Operation Screen (Measurement mode)
(5) When the [CONC] value (concentration) stabilizes near 0.000 and [FLOW] (flowrate) is
0.0, then select [ZERO/SPAN] (by pressing [2]).
Note: [ZERO CALIBRATION]
measures the background level.
(6) Select [START] (by pressing [1]).
→
Background measurement starts (starts counting down six seconds).
Stabilizes near 0.000
Decreases to 0.0
Note: During BG measurement
After 6 sec
Note: BG measurement is completed.