42
8-2 Continuous
Measurements
Continuous measurement refers to manually started and stopped measurements. Default
conditions for continuous measurements are indicated below.
Parameter
Default
Ref. Chap. (page)
Averaging Period
60 min
9-2-2 (58)
Mass Concentration Conversion Factor
1.00
9-2-3 (58)
Dehumidification Heater Temperature
60
°
C
9-2-6 (61)
For more information about default setting values, see 9-2 Settings Menu (page 57).
Starting Measurements
After finishing 8-1-2 Adjusting the Sensitivity Level, display the operation screen (standby mode).
Note: Make sure the scattering plate knob mark is at the [MEASURE] position and the valve
is open.
(1) Select [START] (by pressing [1]).
→
A Confirmation screen is displayed.
Operation Screen (Standby mode)
Confirmation Screen
(2) Select [START] (by pressing [2]).
→
The pump starts operating (the [MEASURE] lamp illuminates) and the measurement
starts.
Confirmation Screen
Operation Screen (Measurement mode)
Note: The [CONC] (dust concentration) and [FLOW] (pump flowrate) values are updated
every second.