(7) Disconnect the cable and two tubes shown.
Note: Disconnect the cable by gripping the
connector on the end of the cable.
Pulling on the cable portion could pull the
cable out of the connector or damage
the connection.
(8) Remove the four screws that fasten the pump
and remove the pump.
Note: Use a Phillips screwdriver (No. 2).
(9) Install a new pump on the pump base.
Note: Reuse the screws removed in step (8).
Note: Use a Phillips screwdriver (No. 2).
(10) Connect the cable from the pump into the
indicated connector and connect the two
tubes as shown.
Tubes
Cable
Four
screws
New
pump
Pump
base
Marked
tube
Cable from pump