Connecting the R&S UP300/350 to the AC Line
R&S UP300/350
E-1147.2759.00
3-42
Operating Manual, 11/2007
3.3
Connecting the R&S UP300/350 to the AC Line
WARNING
Danger of electric shock!
The R&S UP300/350 meets the requirements for Safety Class I according to
DIN EN 61010-1/IEC 61010-1, e.
g. all metal parts that can be touched or
accessed without removing the enclosure are connected to the protective
ground of the power supply network.
When connecting the instrument to the AC power supply, always use a power
cable and a socket with earthing contact.
Automatic AC line
voltage detection
When the R&S UP300/350 is connected to the AC line, it automatically sets
itself to the correct voltage (range: AC voltage 100 V to 240 V, AC frequency
50 Hz to 60 Hz). There is no need to set the voltage manually or change the
fuse.
Connecting the
AC line
1.
Use the supplied power cord to connect the R&S UP300/350 to the AC
line.
The power supply connector [18] is at the rear of the instrument.
2.
Connect the power cord to the AC line.
3.4
Switching On the R&S UP300/350
Note:
The AC line is still connected to the R&S UP300/350 when the
instrument is in the standby mode.
AC line switch on
the rear panel
The R&S UP300/350 is connected to the AC line via power supply connector
[18]. The AC line switch [20] which isolates the R&S UP300/350 from the AC
line is located next to the power supply connector.
ON/STANDBY
switch on the
front panel
ON operating state
After switching on by means of the AC line switch [20] at the rear panel, the
R&S UP300/350 is in standby mode and the yellow LED [2] comes on. If you
press the ON/STANDBY switch [1], the instrument is switched on and the
green LED [2] comes on.
STANDBY operating state
To switch the R&S UP300/350 over from the operating mode to standby
mode, press the ON/STANDBY switch [1] for approx. 2 seconds. After
switching off the ON/STANDBY switch [1], the yellow LED [2] comes on.
Switching on
the R&S UP300/350
1.
Press the AC line switch [20] on the rear panel in the “I” position.
2.
Press the ON/STANDBY switch [20] on the front panel.
The green LED ON [2] comes on.