Generator R&S
UP300/350
E-1147.2759.00
6-134
Operating Manual, 11/2007
Entering the High-Level Time
Use
Use the high-level time (burst duration) to determine the time during which
the sinewave has its high level.
Note:
The setting is always valid for both channels (Ch 1&2).
Entering the
high-level time
1.
Press the
function key
in the
menu.
An entry field containing the currently applicable setting is displayed. The
default setting is “500 ms“. At the same time, the function keys [13] are
assigned various units of measurement.
2.
Enter a new value (
Ê
The permissible entry range is:
0.001 s
≤
HIGH LEVEL TIME
≤
60 s
The new setting is stored and displayed in the parameter field.
Ch 1&2