![Rohde & Schwarz Smart Instruments UP300 Скачать руководство пользователя страница 346](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754346.webp)
System Messages
R&S UP300/350
E-1147.2759.00
8-346
Operating Manual, 11/2007
8 Error
Messages
This chapter
You can find a description of errors that may occur in the R&S UP300/350
and you will also find notes on troubleshooting.
The R&S UP300/350 displays detected errors and warnings on the screen.
The various types of messages are described as the following:
System
messages
Warnings indicating impermissible operating states
Further
information
Chapter 6 describes all the R&S UP300/350‘s menus and the associated
functions in detail.
8.1 System
Messages
System messages
System messages inform you about internally detected errors. The
following information are displayed, e.
g.:
Type of error (
x
)
Four-digit error number (
y
)
Request for closing the system messages (
z
)
The error number allows the service shop to determine the type of error. In
the event of a system message, please write down the error number and
proceed according to the following steps:
Device error
“Error number”
A system error was detected in the instrument.
1.
Please write down the error number and the corresponding instrument
settings.
2.
Contact your nearest Rohde & Schwarz representative (
Ê
instrument may have to be checked in the service shop.
Overtemperature
error
“Error number”
An impermissibly high temperature was detected in the instrument. The
internal fans are switched to full power for approx. 30 seconds, and then the
R&S UP300/350 is automatically switched off to prevent further overheating.
The overtemperature could be caused by too high an ambient temperature
and/or a reduced air circulation.
1.
Allow the instrument to cool off for a while and remove any
obstructions that could prevent the R&S UP300/350 from air circulation.
2.
If this does not eliminate the overtemperature, have the instrument
checked by the service shop.
Note:
Some errors can cause the instrument or parts of the instrument to be
switched off immediately in order to avoid destruction of components.
Whenever a system message occurs, an entry is made under SYSTEM
MESSAGES (
Ê
y
z
x