Generator
Operating Manual, 11/2007
6-143
E-1147.2759.00
Entering the Total Signal RMS
Use
The total voltage is divided between the useful and interference signal in the
(selectable) ratio. The maximum voltage setting is limited by the maximum
peak value. As a result, the limits for the user-selectable RMS depend on the
amplitude ratio.
Selecting
the channel
1.
Press the numeric keys
1
,
2
, or
3
to select channel
Ch 1
,
Ch 2
, or both
channels
Ch 1&2
.
The selected channel is displayed in green in the channel display.
Entering the
total signal RMS
2.
Press the
function key
in the
menu.
An entry field containing the currently applicable setting is displayed. The
default setting is “100 mV (0.1 FS)”. At the same time, the function keys
[13] are assigned various units of measurement.
Analog
Digital
3.
Enter a new value (
Ê
The permissible entry range depends on amplitude ratio:
0
≤
TOTAL RMS
≤
5.299 V
(Analog)
0
≤
TOTAL RMS
≤
0.7063 FS
(Digital)
The new setting is stored and displayed in the parameter field.
(Analog)
(Digital)
Note:
The maximum limit of signal amplitude applies if only the SINE MOD
DIST generator function is set. If other generator functions (e.g. DC OFFSET,
NOISE) are activated, their amplitude will also be taken into
consideration and the maximum signal amplitude decreases accordingly.