Generator R&S
UP300/350
E-1147.2759.00
6-166
Operating Manual, 11/2007
Analog
Digital
4.
Enter a new value (
Ê
The permissible entry range is:
V
Start
≤
Stop
≤
7.495 V
(Analog)
V
Start
≤
Stop
≤
0.9999 FS
(Digital)
within: V
Start
- stop value of amplitude sweep (
Ê
above)
The new setting is stored and displayed in the parameter field.
Analog
Digital
Selecting the Spacing of Sweep Steps (Lin/Log)
Use
The spacing of the sweep points can be selected. You can decide if the dis-
tance between the valus set one after another is in
Linear
or
Logarithmic
steps.
Selecting the
Spacing
1.
Press the
function key
in the corresponding
or
submenu
.
A selection field containing the available settings is displayed. The
default setting is “Linear“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.