![Rohde & Schwarz Smart Instruments UP300 Скачать руководство пользователя страница 345](http://html1.mh-extra.com/html/rohde-and-schwarz/smart-instruments-up300/smart-instruments-up300_operating-manual_1477754345.webp)
R&S UP300/350
Reference Input and Output (10 MHz In/Out)
Operating Manual, 11/2007
7-345
E-1147.2759.00
7.3
Reference Input and Output (10 MHz In/Out)
External
reference
If an external reference is used, the internal reference oscillator is
synchronized to the 10 MHz reference signal at REF IN/OUT connector [23].
The input level must be 0.5 to 2 V.
Internal
reference
The 10 MHz signal from the internal reference oscillator is available at the
REF IN/OUT connector [23] so that other devices can be synchronized to the
R&S UP300/350. The output level is 7 dBm.
Instrument
setting
You can switch over between the internal and external reference in the
CONFIG menu (
Ê
7.4
USB Interface (PC, DEV)
Connector
The USB-Host [16] and USB-Device [17] connectors on the rear panel of the
R&S UP300/350 are for a USB device.
Pin Signal
1 Vbus
(Vcc)
2 D-
3 D+
4 GND
Pin assignment
Shell Shield
7.5
Audio Monitoring Output (MON OUT)
Connector
At the audio monitoring output [15], you can use headphones to monitor
signals which can be tapped at various points in the R&S UP300/350.
Specification
Connector:
Mini jack 3.5 mm
Output impedance:
10
Ω
Voltage:
Max. 2 V (without load)
Instrument setting
The MONITOR menu allows you to make the settings when an external
headphone is used (