Data Sheet
R&S UP300/350
E-1147.2759.00
1-12
Operating Manual, 11/2007
Format
professional and consumer
Measurement functions
RMS value, wideband
Error limits
measurement speed AUTO, at
1 kHz sine, AC coupling
measurement speed AUTO
FAST
±0.1 dB, additional error with
DC coupling
±0.1 % of measurement range
±0.1 dB additional error
Integration time
AUTO FAST/AUTO VALUE
5 ms/50 ms, at least 1 cycle
1 ms to 10 s
Noise with
A
fi
lter, 600
Ω
source
impedance
with CCIR unweighting
fi
lter,
600
Ω
source impedance
< 2 µV
< 4 µV
Filters weighting
fi
lters and sets of prede
fi
ned octave and third-octave
fi
lters; up to 3
fi
lters can be combined
RMS value, selective
Error limits
±0.2 dB
Bandwidth (-3 dB)
Relative
Absolute
1%, 3%, 1/3 octave,
1/12 octave, value
minimum BW 10 Hz
10 Hz to f
max
/5
Selectivity
100
dB
Frequency setting
fi
xed through entered value or
autotuning
Peak value
Measurement
pos. peak, neg. peak,
peak-to-peak, absolute peak
Error limits
at 1 kHz
±0.2 dB
Interval
20 ms to 10 s
Filters weighting
fi
lters and sets of prede
fi
ned octave and third-octave
fi
lters; up to 3
fi
lters can be combined
Quasi-peak
Measurement
in accordance with CCIR 468-4
Error limits
analyzer bandwidth 22 kHz
in accordance with CCIR 468-4
Noise
with CCIR weighting
fi
lter,
600
Ω
source impedance
<12 µV
Filters weighting
fi
lters and sets of prede
fi
ned octave and third-octave
fi
lters; up to 3
fi
lters can be combined