Analyzer R&S
UP300/350
E-1147.2759.00
6-256
Operating Manual, 11/2007
Activating the
averaging
1.
Press the
function key
in the
submenu
.
A selection field containing the available settings is displayed. The
default setting is “OFF“.
2.
Use the
rotary knob
[11] to select a setting.
3.
Press the
ENTER key
[5] to close the selection field.
The new setting is stored and displayed in the parameter field.
Entering the
averaging factor
4.
Press the
function key
in the
submenu
.
An entry field containing the currently applicable setting is displayed. The
default setting is “1“.
5.
Enter a new value (
Ê
The permissible entry range is:
1
≤
AVG FACTOR
≤
256
The new setting is stored and displayed in the parameter field.
Ch 1&2
Ch 1&2